Instruction Manual
8 How to ...?
797 VA Computrace – Software
244
Note: To make sure that always the latest "response curve file"
is taken for the calculation, the name of the "response curve fi-
le" defined for the parameter Response curve on the Determina-
tion
tab (of the window EDIT WORKING METHOD PARAMETERS)
with calibration "RC Sample with response curve", must match
with the name defined for the parameter Sample identifier on
the Determination tab (of the window EDIT WORKING METHOD
PARAMETERS
) with Calibration technique "RC Record response
curve".
Note: To make sure that always the latest "response curve file"
is taken for the calculation, the path of the calibration file defi-
ned for the parameter
Response curve on the Determination tab
(of the window EDIT WORKING METHOD PARAMETERS with
calibration "RC Sample with response curve", must match with
the path defined for the parameter Data folder (for the currently
logged-in user) on the tab
User Directories of the window USER
RIGHTS
.
Sample table
In the Sample table, list (according to the arrangement on the sam-
ple rack) for every empty vessel on the sample rack a method with
Calibration technique RC Record response curve and for every sam-
ple a method with Calibration technique RC Sample with response
curve: