8 How to ...?
797 VA Computrace – Software
242
Es wird für beide Dosino ein "Prep-Zyklus" empfohlen.
On the Automation tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <Default>).
Method Parameters for the 797
Two steps are required for Suppressor determination. One to re-
cord the response curve with Calibration technique "RC Record re-
sponse curve", another for the actual sample determination with
Calibration technique "RC Sample with response curve". Set CVS or
CPVS for Mode for both methods.
Recording of the Response Curve::
Choose "RC Record response curve" for
Calibration.
Settings in the Dosinos window:
After the addition of Electrolyte solution, Suppressor standard
solution is added via Dosino 3. That should be specified on the
Substances tab of the EDIT WORKING METHOD PARAMETERS
window (here the window for CVS):