8 How to ...?
797 VA Computrace – Software
230
50 mL sample vessels can be placed on the two outer rings (11 mL
sample vessels on the two inner rings).
Arrangement: Place on the outer of the two rings alternately "In-
tercept solution" and sample solution, on the inner of the two rings
the rinsing solution.
Example for an arrangement with 7 samples:
Intercept solution
Rinsing solution
not used
Sample
1
2
3
28
27
29
30
31
56
55
57
58
59
84
83
85
86
87
112
111
Note: Although the first sample is placed on position 2, position
1 remains the starting position in the start sequence of the
method at the 838 (recording the "Intercept value").
Note: Before each start, define the position of the first sample
vessel for the 838 parameter "SAMPLE" with the 838 keypad.
General Settings in the 797 Software
Activate both options on the
General tab of the GENERAL SET-
TINGS
window, so that a recorded "Intercept value" replaces the
old one automatically.
Settings on the Dosinos tab of the GENERAL SETTINGS window in
the 797 Software: