Instruction Manual

8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
229
Instruments
Install the 838 Advanced Sample Processor, a 800 Dosino, a
732 Relay Box with two 823 Membrane Pump Units (see
Hardware Manual 797 and 838 Instructions for Use).
Dosinos:
(Dosino 1: not used)
Dosino 2: 2 mL Exchange Unit Brightener standard so-
lution
(Dosino 3: not used)
Method at 838
Before each start, define the position of the first sample vessel for
the 838 parameter "SAMPLE" with the 838 keypad.
Set method LAT at the 838 Advanced Sample Processor.
LAT
parameters
method LAT
number of samples rack > Number of samples to be
processed (entire sample
rack)
>start sequence
1 CTL:Rm: INIT > Initialize Remote lines
2 Move 1 : sample > First Sample on Pos. "SAM-
PLE"
3 CTL:Rm: ************1 > Remote start CT797
4 CTL:Rm: ************0
>sample sequence
1 SCN:Rm: *****1** > Scan line for incoming signal
from CT797
2 MOVE 1 : sample > Move needle to sample
3 LIFT: 1 : work mm > Place needle at working posi-
tion
4 PERISTALT: 300 s 10 > Sample transfer
5 CTL:Rm: ************1 > Set line: sample transfer
ready
6 CTL:Rm: ************0
7 MOVE 1 : +28 > Move needle to rinsing posi-
tion
8 LIFT: 1 work mm > Place needle at working posi-
tion
9 PERISTALT: 5 s 10 > Suck rinsing solution in the
needle
10 SCN:Rm: *****1** > Scan line for incoming signal
from CT797
11 PERISTALT: 300 s 10 > Rinse needle
12 CTL:Rm: ************1 > Set line: needle cleaning
ready
13 CTL:Rm: ************0
Samples