Instruction Manual
8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
227
General Settings in the 797 Software
Settings on the Dosinos tab of the GENERAL SETTINGS window in
the 797 Software:
It is recommended to do for all used Dosinos one "Prep cycle".
On the Automation tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <Default>).
Method Parameters for the 797
Set CVS or CPVS for
Mode, and "MLAT" for Calibration.
Settings for the
DOSINOS window:
Checking Use for predose for Dosino 1, VMS and additional Sup-
pressor concentrate are added to the measuring vessel before the
sample transfer. Together, they form the "Intercept solution". To
keep the Suppressor concentration stable, additional Suppressor
concentrate is added after the sample transfer.