Instruction Manual

8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
225
LAT
parameters
method LAT
number of samples rack > Number of samples to be
processed (entire sample
rack)
>start sequence
1 CTL:Rm: INIT > Initialize Remote lines
2 Move 1 : sample > First Sample on Pos. "SAM-
PLE"
3 CTL:Rm: ************1 > Remote start CT797
4 CTL:Rm: ************0
>sample sequence
1 SCN:Rm: *****1** > Scan line for incoming signal
from CT797
2 MOVE 1 : sample > Move needle to sample
3 LIFT: 1 : work mm > Place needle at working posi-
tion
4 PERISTALT: 300 s 10 > Sample transfer
5 CTL:Rm: ************1 > Set line: sample transfer
ready
6 CTL:Rm: ************0
7 MOVE 1 : +28 > Move needle to rinsing posi-
tion
8 LIFT: 1 work mm > Place needle at working posi-
tion
9 PERISTALT: 5 s 10 > Suck rinsing solution in the
needle
10 SCN:Rm: *****1** > Scan line for incoming signal
from CT797
11 PERISTALT: 300 s 10 > Rinse needle
12 CTL:Rm: ************1 > Set line: needle cleaning
ready
13 CTL:Rm: ************0
Samples
With 50 mL sample vessels the two outer rings are used, with 11
mL sample vessels the two inner rings.
Arrangement: Place samples on the outer of the two rings, place
rinsing solutions on the inner of the two rings.
Example for an arrangement with 11 samples:
With 50 mL sample vessels: first sample on position 1, first
rinsing solution on position 29.