Instruction Manual
8 How to ...?
797 VA Computrace – Software
218
13. If you want to save the modified determination under the same
name, click on
or MAIN WINDOW / File / Save determina-
tion
. The old file will be overwritten.
14. If you want to save the modified determination under a new
name, click on MAIN WINDOW / File / Save determination as.
Select the desired directory, enter the determination file name
*.dth in the SAVE AS window, and click the <Save> button.
Print determination results and curves
1. Click on or MAIN WINDOW / File / Print. The PRINT OP-
TIONS
window is opened.
2. Select the elements, which should be printed (see Printing in
determination mode, section 5.7).
3. Select the order of printout for each element checked.
4. Close the PRINT OPTIONS window by clicking the <OK> button.
8.6 Analyze Electroplating Bath Solutions
Introduction
The 797 VA Computrace can perform electroplating bath analysis.
Two modes are especially designed for Electroplating Bath VA: CVS
(Cyclic Voltammetric Stripping) and CPVS (Cyclic Pulse Voltammet-
ric Stripping).
With these two modes, 8 different Calibration techniques can be
chosen:
For method optimization:
“Standard addition plating bath” (this Calibration
technique is used to develop and optimize methods)
For Brightener analyzing:
“MLAT Standard addition for brighteners” (Modi-
fied Linear Approximation Technique)
“LAT Standard addition for brighteners” (Linear
Approximation Technique)
"LAT Record intercept value” (is used to record the
"Intercept value")
For Suppressor analyzing ("Dilution Titration Technique"):
"Dilution Titration Technique"
“DT Record calibration curve“ (used to record the
calibration curve)