Instruction Manual

8 How to ...?
797 VA Computrace – Software
214
15. Click the <OK> button. The determination is saved automati-
cally if specified on the
General tab of the GENERAL SETTINGS
window and the result report is printed if specified on the Do-
cumentation
tab of the EDIT WORKING METHOD PARAMETERS
window.
Perform determinations using the 863 Compact VA Autosampler
1. Install the 863 Compact VA Autosampler (see Installation of
863 Compact VA Autosampler, section 1.3) and set method
2.
2. Click on MAIN WINDOW / Settings / General settings and select
the Hardware tab.
3. Define the Sample processor field of the Hardware tab.
4. Modify the parameters for Automation in the Automation tab
of the GENERAL SETTINGS window as desired.
5. If desired, test the automation parameters: Click on MAIN WIN-
DOW
/ Settings / General settings and select the Automation
tab. Fill two sample vessels with water and place them on the
sample rack. Click on , check the automation pa-
rameters and modify them if needed.
6. Click on or MAIN WINDOW / Mode / Determination.
7. Click on or MAIN WINDOW / Window / Working method
specification
to open the WORKING METHOD SPECIFICATIONS
window.
8. Load the desired method into the WORKING METHOD SPECIFI-
CATIONS
window (see How to Load a method, section 8.4).
9. If desired, modify and save the loaded method (see How to
Edit the working method, section 8.4).
10. Click on
or MAIN WINDOW / Window / Sample table to
open the SAMPLE TABLE window.
11. Load the desired sample table or edit the current sample table
(see Sample table, section
5.6).
12. Transfer the desired sample amount into the sample vessels.
Place the sample vessels at the odd positions on the sample
rack of the 863 Compact VA Autosampler. For each sample
vessel, place a vessel filled with rinsing solution at the follow-
ing even position (volume rinsing solution = volume sample so-
lution).
13. Click on or MAIN WINDOW / Window / Monitor to open the
MONITOR window.
14. Start the measurement by clicking the
icon in the MAIN
WINDOW
or the button in the MONITORING window.