8 How to ...?
797 VA Computrace – Software
210
Copy parameters from determination methods
1. Click on or MAIN WINDOW / Mode / Determination.
2. Click on or MAIN WINDOW / File / Load determination.
3. Select the desired determination file *.dth in the OPEN window
and click <OK>. The determination is loaded into the DETERMI-
NATION CURVES
window.
4. Click on DETERMINATION CURVES / Edit / Copy parameters to
working method
.
Copy parameters from signal files
1. Click on or MAIN WINDOW / Mode / Exploratory.
2. Click on or EXPLORATORY SPECIFICATION / File / Load
signal
.
3. Select the desired signal file *.sig in the OPEN window and click
<OK>.
4. Click on EXPLORATORY SPECIFICATION / Transfer / Parameters
/ To w
orking method.
Save the working method
1. If you want to save a modified working method under the
same name, click on or MAIN WINDOW / File / Save
method
. The old file will be overwritten.
2. If you want to save the working method under a new name,
click on
MAIN WINDOW / File / Save method as. Select the de-
sired directory, enter the method file name *.mth in the SAVE
AS
window, and click the <Save> button.
Edit the working method
1. Click on or MAIN WINDOW / Mode / Determination.
2. Click on or MAIN WINDOW / Window / Working method
specification
to open the WORKING METHOD SPECIFICATIONS
window.
3. Click on or MAIN WINDOW / File / Load method.
4. Select the desired method file *.mth in the OPEN window and
click <OK>. The method is loaded into the WORKING METHOD
SPECIFICATIONS
window.
5. Set the parameters in the
WORKING METHOD SPECIFICATIONS
window
to the desired values (see Working method specifica-
tions window, section 5.2).