Instruction Manual

6 Electroplating Bath VA
797 VA Computrace – Software
184
Sample identifier [ 32 characters ; "sample" ]
Identification for sample.
Note: The Sample identifier is used for the file na-
me. To make sure that for "Suppressor analysis
with 838 Advanced Sample Processor and RC" al-
ways the latest response curve file is taken for the
calculation, this Sample identifier must match with
the name of the calibration file defined for the pa-
rameter Response curve on the Determination tab
with calibration "RC Record response curve".
Example for Sample identifier:
Response Suppressor
Note: To make sure that for "Suppressor analysis
with 838 Advanced Sample Processor and RC" al-
ways the latest response curve file is taken for the
calculation, the path of the calibration file defined
for the parameter Response curve on the Sub-
stances
tab with Calibration technique "RC Sample
with response curve”, must match with the path
defined for the parameter Data folder (for the cur-
rently logged-in user) on the tab User Directories
of the window USER RIGHTS.
Note: A determined response curve file is only
adopted automatically if the checkbox Save cali-
bration curves additionally without date and time
on the General tab of the GENERAL SETTINGS
window is checked.
Volume electrolyte (mL) [ > 0 mL ; 10 mL]
The volume of the Electrolyte Solution (see
Electrolyte solution, section 6.4).
Cell volume (mL) [ > 0 mL ; 10 mL ] [ read only ]
Total volume of solution (equal to the volume of
the Electrolyte solution) in the measuring vessel at
the start of the determination. The sample con-
centrations Conc. calculated refers to this cell vol-
ume.
Initial electrode conditioning (see Initial electrode conditioning,
section 6.3)
Check the checkbox to use conditioning.
No. of conditioning measurements [ 1...100 ; 1 ]
Number of replications of conditioning measure-
ments. The “total number of conditioning mea-
surements” is the
No. of conditioning measu-
rements
multiplied with the number of Save last ..
sweeps
(Voltammetric tab). “Total number of con-
ditioning measurements” must not exceed 100.