Instruction Manual
6 Electroplating Bath VA
797 VA Computrace – Software
176
cations multiplied with the number of Save last ..
sweeps
(Voltammetric tab). “Total number of
measurements” must not exceed 10.
Voltammetric tab with Calibration technique “DT Record
calibration curve”
The Voltammetric tab of the EDIT WORKING METHOD PARAME-
TERS
window contains parameters for preparation procedures and
VA measurement modes. With the Calibration technique "DT Re-
cord calibration curve" the Voltammetric parameters are not edit-
able. They depend on the parameters defined for recording the
calibration curve.
Following parameters have to be defined:
Initial mixing time (s) [ read only ] [ 0...80600 s ; 10 s ](see Initial
mixing time with CVS and CPVS, section 6.3)
Time of mixing before the first measurement of
the sample solution.
Conditioning Cycles [ read only ] (see Conditioning cycles with CVS
and CPVS, section 6.3)
Start potential (V) [ -5...+5 V ; 0.2 V ] [ read only ]
Start voltage for the cyclic conditioning sweep.
End potential (V) [ -5...+5 V ; 1.625 V ] [ read only ]
Final voltage for the cyclic conditioning sweep.
No. of cycles [ 0...X ; 0 ] [ read only ]
Number of conditioning cycles.
Pretreatment (see Pretreatment with CVS and CPVS, section 6.3)
Cleaning potential (V) [ -5...+5 V ; 1.625 V ] [ read only ]
Voltage applied to the electrodes during the
Cleaning time.
Cleaning time (s) [ 0...80600 s ; 0 s ] [ read only ]
Time during which the Cleaning potential is app-
lied to the electrodes.
Equilibration potential (V) [ -5...+5 V ; 1.625 V ] (This pa-
rameter is with CPVS part of the sweep)
Voltage applied to the electrodes during the
Equilibration time.
Equilibration time (s) [ 0...80600 s ; 5 s ] (This parameter is
with CPVS part of the sweep)
Waiting time before starting the sweep; with Equi-
libration potential
applied to the electrodes. If the
checkbox
Hydrodynamic (measurement) is not che-
cked, the stirrer is switched off during that time.
Sweep [ read only ]
The sweep parameters depend on the mode (see
CVS - Cyclic Voltammetric Stripping section 3.2,