Instruction Manual
6 Electroplating Bath VA
797 VA Computrace – Software
170
Note: If both No. of conditioning measurements [
1...100 ; 1 ] and Auto Std.dev (%) [ >0.1 ; 1 ] are
activated, the conditioning is finished as soon as
the first of the two criterias is reached.
Addition mixing time [ 0...80600 s ; 10 s ]
Time of mixing after Additions (for the first mea-
surement, the
Initial mixing time is used).
Calibration curve [ path + file name ; ]
Define the calibration curve file (the file in which
the determination of the calibration was(will be)
saved). The calibration curve can be recorded with
the
Calibration technique “DT Record calibration
curve”.
Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the name of the calibration file defined
for Calibration curve, must match with the para-
meter Sample identifier on the Determination tab
with Calibration technique "DT Record calibration
curve".
Example for Calibration curve:
C:\User XYZ\Data\Calibration Suppressor.dth
Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the path of the calibration file defined for
Calibration curve, must match with the path de-
fined for the parameter
Data folder (for the cur-
rently logged-in user) on the tab User Directories
of the window USER RIGHTS.
Note: A determined calibration curve is only
adopted automatically if the checkbox Save cali-
bration curves additionally without date and time
on the General tab of the GENERAL SETTINGS
window is checked.
No. of replications [ 1...10 ; 2 ]
Number of replications (= total number of measu-
rements) for each variation (sample, standard ad-
dition, calibration level, conditioning measure-
ment). For cyclic modes (CV, CVS, CPVS) the “total
number of measurements” is the
No. of repli-
cations
multiplied with the number of Save last ..