Instruction Manual

6 Electroplating Bath VA
797 VA Computrace – Software
162
SPECIFICATIONS window. With “LAT Standard addition for brighte-
ners” following parameters have to be defined:
Substance [ 24 characters ; ]
Substance name. For the assignment of a found
peak to this substance the checkbox on the left
side of the substance name must be checked.
With CVS: Peak pos. +/- (V) [ -5...+5 V ; 0 V ]
Position of the peak voltage for the substance and
tolerance for this verification voltage.
With CPVS: Step potential [ read only ; step 1: 0.2 V ]
Displays the potential values, which are defined in
the window edit stripping steps.
Bsln.
Parameters for baseline evaluation (details see
Baseline, section 5.2). Click the button to open
the BASELINE window for the selected substance.
Additive solution
Definition of addition solutions for standard addi-
tion.
No. [ 0...8 ; 0 ]
Number of addition solution used for manual or
automatic addition. For automatic additions, this
number is also the Dosing Device number. If a
mixed standard solution is used, the number of
this solution must be entered for each substance
present in the mixed solution.
Achtung: Dosino 1..3 refer to Dosinos connected
to the 797 VA Computrace. Dosino 4...7 refer to
Dosinos connected to the 846 Dosing Interface (to
MSB 1...4).
Conc. [ > 0 ; 0 ]
Value for concentration of addition solution.
Unit [ fL/L...mL/L ; mL/L ]
Unit for concentration of addition solution.
Volume (mL) [ > 0.01 mL / var ; 0 mL ]
Addition volume. For entering variable addition
volumes, click the button to open the EDIT
VARIED ADDITION
window for the selected sub-
stance (details see Variable addition, section 5.2).
In this case, var is entered into the field instead of
a fixed value. This field only appears once for solu-
tions with the same number (mixed standards)
and it is not displayed if 0 is entered as solution
number.