Instruction Manual

6 Electroplating Bath VA
797 VA Computrace – Software
160
on the General tab of the GENERAL SETTINGS
window is checked.
Initial electrode conditioning (see Initial electrode conditioning,
section 6.3)
Check the checkbox to use conditioning.
No. of conditioning measurements [ 1...100 ; 1 ]
Number of replications of conditioning measure-
ments. The “total number of conditioning meas-
urements” is the No. of conditioning measure-
ments
multiplied with the number of Save last ..
sweeps
(Voltammetric tab). “Total number of con-
ditioning measurements” must not exceed 100.
or Auto Std.dev (%)[ >0.1 ; 1 ]
Define until which „Std.dev.“ the conditioning is
continued. The „Std.dev.“ is calculated from all sa-
ved sweeps of the last two “conditioning meas-
urements”.
Note: If you set a small value (e.g. < 1%), it can
take a long time until reaching the target.
Note: If both No. of conditioning measurements [
1...100 ; 1 ] and Auto Std.dev (%) [ >0.1 ; 1 ] are
activated, the conditioning is finished as soon as
the first of the two criterias is reached.
Addition mixing time [ 0...80600 s ; 10 s ]
Time of mixing after Additions (for the first meas-
urement, the
Initial mixing time is used).
No. of additions [ 0...28 ; 2 ]
Number of additions of additive standard solution.
No. of replications [ 1...10 ; 2 ]
Number of replications (= total number of measu-
rements) for each variation (sample, standard ad-
dition, calibration level, conditioning measure-
ment). For cyclic modes (CV, CVS, CPVS) the “total
number of measurements” is the No. of replica-
tions
multiplied with the number of Save last ..
sweeps
(Voltammetric tab). “Total number of
measurements” must not exceed 10.
Voltammetric tab with Calibration technique “LAT”
The Voltammetric tab of the EDIT WORKING METHOD PARAME-
TERS
window contains parameters for preparation procedures and