Instruction Manual
5.8 Data processing and evaluation
797 VA Computrace – Software
141
mined by the baseline parameters set for this substance in the
BASELINE window (see Baseline, section 5.2). The following possi-
bilities exist for baseline calculation:
Linear
Front end
Linear
Rear end
Linear
Whole Peak
Start End
Exponential
Whole Peak
Start End
Polynomial
Whole Peak
Start End
Start End
Horizontal
Whole Peak
As default, the start and end base points of the baselines are calcu-
lated automatically. They can be set to fixed voltage values if de-
sired. The baseline scope Front end or Rear end for linear baselines
should be selected for asymmetric or double peaks.
Evaluation quantity calculation
The peak evaluation quantity is identical for all peaks of a determi-
nation and must be set on the Substances tab of the EDIT WOR-
KING METHOD PARAMETERS
window (see Working method speci-
fications window, section 5.2). With the aid of the calculated base-
lines, the set evaluation quantity
Height, Area, Derivative or Coulo-
metric
is determined for each substance peak and displayed as a
result.