Instruction Manual

5.6 Sample table
797 VA Computrace – Software
135
Sample ID [ 32 characters ; "sample" ]
Identification for sample.
Amount [ > 0 ; 10 ]
Amount of sample added to the measuring vessel.
Unit [ mL, g ; mL ]
Selection of unit for sample amount.
Cell volume (mL) [ > 0 mL ; 10 mL ]
Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer) in the measuring vessel at the
start of the determination. The sample concentra-
tions Conc. refer to this cell volume.
Method
Selection of the method used for this sample.
Status [ read only ]
Display of sample determination status: Measur-
ing
, Done or "empty" (ready for start).
Add a new sample data row. The SAMPLE win-
dow for entry of Method, Sample ID, Sample
amount
, Sample unit, and Cell volume is opened.
Edit the selected sample data row. The SAMPLE
window for modification of Method, Sample ID,
Sample amount, Sample unit, and Cell volume is
opened.
Reset the Status of the sample data rows to "emp-
ty" in order to restart the current sample table.
Clear the contents of the selected sample data
row.
Print the sample table.
Load/save sample table
Sample table files (
*.spt) containing sample table data can be cre-
ated, loaded and saved by the following commands:
SAMPLE TABLE / File / New
Load an empty sample table into the working
memory.
SAMPLE TABLE / File / Load
Load an existing sample table into the working
memory. The name of the sample table is dis-
played in the status bar of the
MAIN WINDOW.
SAMPLE TABLE / File / Save
Save the current sample table loaded in the wor-
king memory. If the sample table has been chan-
ged since loading, the message The file already
exists. Overwrite?
appears. Click Yes to overwrite