Instruction Manual

5.4 Determination curves
797 VA Computrace – Software
121
List of curves
The top subwindow lists all available curves of the determination
with the evaluated peak heights or peak charges .
Determination curves
The lower left subwindow shows a single or all determination
curves.
Calibration curves
At the right side of the determination curves, there are eight sub-
windows for display of each substance calibration curve.
Edit determination method parameters
DETERMINATION CURVES / Edit /
Determination method parameters
Open the EDIT DETERMINATION METHOD PARA-
METERS
window for viewing and modifying the
parameters of the determination method used to
record the determination. This window contains
the tabs Specifications, Determination, Voltammet-
ric
, Substances and Calculations. If parameters in
the EDIT DETERMINATION METHOD PARAMETERS
window are changed and <OK> is pressed, the de-
termination will be recalculated automatically.
Note: Irrespective whether fields have been modified or not, the
modification date is entered into the Modified line if the EDIT DE-
TERMINATION METHOD PARAMETER
window is closed by clicking
<OK>. If the determination shall not be marked as “modified”, the
EDIT DETERMINATION METHOD PARAMETER window has to be
left with <Cancel>.
Note: These parameters will not automatically be used for the
next measurement. To do this, use the Copy parameters to working
method
command (see Copy parameters to working method, sec-
tion 5.4).
Specifications
The Specifications tab of the EDIT DETERMINATION METHOD PA-
RAMETERS
window contains the main specifications of the deter-
mination method. Only
Title, Remark1 and Remark2 can be edited.