Instruction Manual
5.3 Monitor
797 VA Computrace – Software
117
Calibration curve id [ 32 characters ; "sample" ]
Identification for calibration curve.
Add the calibration solution into the measuring
vessel and press <OK>.
MANUAL ADDITION
This window appears at the start of each manual
addition of standard solutions for standard addi-
tion determinations or the recording of calibration
curves.
Add (from solution no. X) (ml) [ > 0.01 ml ; 0 ml ]
The addition volume for the manual addition de-
fined in the Substances tab is displayed and can be
changed if desired. Add the addition solution into
the measuring vessel and press <OK>.
BATCH SOLUTION EXCHANGE
This window appears if the Batch with solution ex-
change
option is selected for the measurement
technique. The number of the solution to be pla-
ced and the substances and their concentration
defined in the Substances tab are displayed. Add
this solution into the measuring vessel and press
<OK>.
END OF DETERMINATION
This window appears at the end of the determina-
tion. By pressing
<OK>, the determination is
automatically saved if the
Auto save determination
and signal
option is enabled in the General tab of
the GENERAL SETTINGS window. The report ele-
ments checked in the Documentation tab are au-
tomatically printed out.
Graphical properties for monitoring curves
The graphical properties for curves in the
MONITORING window
can be set by selecting the options of the context sensitive menu.
MONITORING / Page properties
The page properties of the MONITORING window
can be set with the page tab of the GRAPHICAL
PROPERTIES
window (details see Page properties,
section 3.5).
The properties of the x and y axis can be set with
the x axis and y axis tab of the GRAPHICAL PROP-