Instruction Manual

5 Determination mode
797 VA Computrace – Software
94
For a detailed description of these parameters see section 3.4 Gen-
eral operation sequence, and section 3.2 VA measurement modes.
If you work with the electroplating bath modes CVS or CPVS, see
section 8.6 Operate a sequence in Electroplating Bath VA.
Note: For Calibration technique "Sample with calibration curve"
the Voltammetric tab parameters are not editable. To make sure
that the Voltammetric parameters defined for the determination of
the sample match with the Voltammetric parameters defined for re-
cording the calibration curve (with Calibration technique "Record
Calibration curve"), following procedure is recommended:
1. Load the determination of the calibration curve (
FILE / Load de-
termination
)
2. Open window
DETERMINATION CURVES
3. Copy the parameters to the working method (EDIT / Copy pa-
rameters to working method
)
The same procedure is recommended for the Suppressor determi-
nation (see Voltammetric tab with Calibration technique "DT Supp-
ressors with calibration curve").
Substances
The Substances tab of the EDIT WORKING METHOD PARAMETERS
window contains parameters for the definition and recognition of
substances, for the definition of standard solutions, for peak
evaluation and results calculation. The parameters displayed de-