Manual

3.2 Remote control commands
794 Basic Titrino
95
&Info
Object Description Input range Reference
& Root
:
Info Current data
: .Report Transmission of formatted reports $G 3.2.2.67.
.Select Report type configuration, parameters,
smpl data, statistics, silo, calib
C-fmla, def, user method, full,
short, mplist, curve, derive, comb,
scalc full, scalc srt, calc,
all, ff ditto
.CalibrationData pH calibration data $G 3.2.2.68.
.Inp1 For measuring input 1
.pHas Asymmetry pH 0...7.00...±20.00 ditto
.Slope Slope of electrode 0...1.000...±9.999 ditto
.Temp Calibration temperature -170.0...25.0...500.0 ditto
.Date Date of calibration read only ditto
.ElectrodeId Id of the calibrated electrode read only ditto
.Inp2 For measuring input 2, as for input 1
.Diff For differential input, as for input 1
.Checksums Checksums $G 3.2.2.69.
.MPList Checksum of meas.point list read only ditto
.ActualMethod Checksum of current method read only ditto
.DetermData Determination data $G 3.2.2.70.
.Write Read/write for several nods ON, OFF
.ExV Volume of Exchange/Dosing unit read only/read+write ditto
.MPList Measuring point list
.1 Measuring point 1
.Attribute Attribute read only/read+write ditto
.X X coordinate read only/read+write ditto
.Y Y coordinate read only/read+write ditto
for each measuring point
.TitrResults Titration results
.RS Calculated results
.1 1
st
result
.Value Value read only 3.2.2.71.
up to 9 results
"Info", continuation
.EP Endpoint
.1 1
st
result
.V Value read only ditto
.Meas Measured value read only
.Mark Mark if more than 1 EP per window read only
up to 9 EP's
.Var Variables C4X
.C40 Start measured value read only/read+write ditto
.C41 Titration end volume read only/read+write
.C42 Titration time read only/read+write
.C43 Volume drift in SET read only/read+write
.C44 Titration temperature read only/read+write