Manual

3.2. Remote control commands
794 Basic Titrino
118
3.2.2.79. Info.ActualInfo.Titrator.CyclNo read only
Info.ActualInfo.Titrator.V read only
Info.ActualInfo.Titrator.Meas read only
Info.ActualInfo.Titrator.dVdt read only
Info.ActualInfo.Titrator.dMeasdt read only
Info.ActualInfo.Titrator.dMeasdV read only
Info.ActualInfo.Titrator.ERC read only
$Q sends the current values in the following formats:
DET MET SET MEAS CAL
CyclNo 127 127 127 127 127
V(ml) 1.2345 1.2345 1.2345 - -
Meas:
pH
U, Ipol (mV)
Upol (uA)
T (°C)
3.345
-345.6
-12.5
-
3.345
-345.6
-12.5
-
3.6(mV)
-345.6
-12.5
-
3.345
-345.6
-12.5
25.0
3.345
-
-
-
dVdt (ul/s) - - 2.5142 - -
dMeasdt
pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
0.7957
0.7957
-
-
dMeasdV (mV/ul) - - 10.6326 - -
ERC 34 - - - -
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.82). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.80. Info.ActualInfo.MeasPt.Index read only
Info.ActualInfo.MeasPt.X read only
Info.ActualInfo.MeasPt.Y read only
Info.ActualInfo.EP.Index read only
Info.ActualInfo.EP.X read only
Info.ActualInfo.EP.Y read only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with DET, MET.
.MeasPt.X"165" Volume (DET, MET)
.MeasPt.Y"3.654" Measured value (DET, MET)
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.81. Info.ActualInfo.Display.L1 up to 24 ASCII characters
Info.ActualInfo.Display.L2 up to 24 ASCII characters
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
Lock the display, see 3.2.2.91.
$Q sends the contents of the corresponding display line.