User guide

3 Operation
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
24
3.2 Manual operation
The 793 IC Sample Prep Module cannot be manually operated unless
the installation has been carried out properly as described in section 2
(mains connection, attaching the pump tubing, connection to the IC
system).
Switch instrument on/off
POWER
The 793 IC Sample Prep Module is switched on and off using
mains switch 13
1313
13 on the rear of the instrument (see Fig. 2):
I Instrument switched on
0 Instrument switched off
After the instrument has been switched on the power lamp 1
11
1
lights up to show that the instrument is ready for use.
Switch off remote control of the pump
ON
OFF
REM
TE
In order to allow manual operation the remote control of the
pump must be switched off with switch 2
22
2.
ON Remote control switched on
OFF Remote control switched off
Switch pump on/off
ON
OFF
PUMP
The drive of the pump is switched on and off with the switch 3
33
3
<PUMP>:
ON Pump switched on
OFF Pump switched off
Switch off remote control of the SP Module A
ON
OFF
REM
TE
In order to allow manual operation the remote control of the SP
Module A must be switched off with switch 6
66
6.
ON Remote control switched on
OFF Remote control switched off
Rotate SP Module A
STEP
ACTUATOR
The SP Module A is switched to the next position by means of
switch 8
88
8 <STEP>.
The position display 7
77
7 beside the switch shows whether the SP
Module A is in the correct position:
Lamp on Right position
Lamp off Wrong position or blocked
(procedure see section 4.1)