793 IC Sample Prep Module 793 IC Sample Prep Module Metrohm POWER CH-9101 Herisau/Schweiz E-Mail info@metrohm.com Internet www.metrohm.com REMOTE PUMP REMOTE ON ON ON OFF OFF OFF ACTUATOR STEP Instructions for Use 8.793.
CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm.com 793 IC Sample Prep Module Instructions for Use 8.793.1003 06.
Table of contents Table of contents 1 Introduction............................................................................................ 1 1.1 Instrument description ............................................................................ 1 1.2 Parts and controls .................................................................................... 2 1.3 Information about the Instructions for Use ......................................... 4 1.3.1 Organization....................................
Table of contents 5 Appendix ................................................................................................. 38 5.1 Technical data......................................................................................... 38 5.2 Standard equipment .............................................................................. 40 5.3 Optional accessories ............................................................................. 41 5.4 Warranty and conformity ........................
1.1 Instrument description 1 Introduction 1.1 Instrument description The 793 IC Sample Prep Module consists of a reactor block (SP Module A) with its corresponding control unit; this contains an additional 2-channel peristaltic pump. Inline sample pretreatment like neutralization or cation exchange for ion chromatographic analyses can be carried out with the cation exchanger built into the SP Module A.
1 Introduction 1.2 Parts and controls 1 2 3 4 5 6 793 IC Sample Prep Module 7 8 Metrohm POWER REMOTE PUMP REMOTE ON ON ON OFF OFF OFF 12 11 10 ACTUATOR STEP 9 Fig.
1.2 Parts and controls 13 14 Type 110V-120V 0,5A(T) 220V-240V 0,25A(T) 50 - 60 Hz 50 VA Remote Breakpoint 't WARNING - Fire Hazard - For continued protection replace only with the same type and rating of fuse 20 Actuator Actuator C A S 19 Pump B Made by Metrohm Herisau Switzerland 18 17 16 15 Fig.
1 Introduction SP Module A Ion analysis 1 2 21 21 Reactor block 22 Actuator connection with permanently fixed inlet and outlet capillaries 23 Connection cable Connection cable to control unit 3 22 23 Fig. 3: 1.793.0110 SP Module A 1.3 Information about the Instructions for Use Please read through these Instructions for Use carefully before you put the 793 IC Sample Prep Module into operation.
1.3 Information about the Instructions for Use 1.3.2 Notation and pictograms The following notations and pictograms (symbols) are used in these Instructions for Use: Switch or key 15 Part or control of 793 89 Part or control of 732/733 "Range" Parameter or entry value at 732 IC Detector >PARAM/detector range: 1.
1 Introduction 1.4 Safety notes 1.4.1 Electrical safety While electrical safety in the handling of the 793 IC Sample Prep Module is assured in the context of the specifications IEC 1010-1 (protection class I, degree of protection IP20), the following points should be noted: x Mains connection Setting the mains voltage, checking the mains fuse and the mains connection must be effected in accordance with the instructions in section 2.2.
2.1 Setting up the instrument 2 Installation 2.1 Setting up the instrument 2.1.1 Packaging The 793 IC Sample Prep Module is supplied together with the separately packed accessories in special packagings containing shockabsorbing foam linings designed to provide excellent protection. The actual instrument is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they can assure damage-free transport of the instrument. 2.1.
2 Installation 2.2 Mains connection Follow the instructions below for connecting to the power supply. If the instrument is operated with the mains voltage set wrongly and/or wrong mains fuse there is a danger of fire! 2.2.1 Setting the mains voltage Before switching on the 793 IC Sample Preparation Module for the first time, check that the mains voltage set on the instrument (see Fig. 2) matches the local mains voltage.
2.2 Mains connection 220 – 240 V 100 – 120 V 13 13 Mains switch 19 19 Mains connection plug 20 20 Fuse holder 220 - 240 V 100 - 120 V 220 - 240 V 100 - 120 V Fig. 4: Setting the mains voltage 2.2.2 Fuses One of the two fuses 0.5 A/slow-blow for 100}120 V or 0.25 A/slowblow for 220}240 V is installed in fuse holder 20 of the 793 IC Sample Prep Module as standard.
2 Installation 2.3 Connection to a Modular IC System for Neutralization An important application for the 793 IC Sample Prep Module is neutralization of an alkaline sample for anion determination by IC with chemical suppression. Strong bases (e.g. 30 % NaOH) should not be led unlimited to the SP Module A. To provide a sufficient capacity for exchanging Na+ by H+ Ions, only a small fraction (e.g. 20 µL) is directed to the SP Module A using a sample loop.
2.3 Connection to a Modular IC System for Neutralization Please note, that the suppressor of the 753 IC Suppressor Module is not switched by the 766 Sample Processor program. The remote connection 16 (suppressor/actuator) of the 753 IC Suppressor Module is connected to the event line “fill” of the 733 valve B instead, using the 6.2128.180 cable. Thus the suppressor is switched when the fill position is accessed by valve B. 2.3.
2 Installation The inlet and outlet capillaries mounted on the SP Module A are then connected up. Proceed as follows: 1 Insert SP Module A x Place SP Module A 21 on the floor of the inner chamber of the 733 Separation Center. 2 Connect SP Module A x Remove the plastic stopper from opening 41 or 43 on the 733 IC Separation Center and push cable 23 mounted on the SP Module A 21 through the opening. x Connect cable 23 to connection 17 “Actuator“ of the 793 IC Sample Prep Module (see Fig. 5).
2.3 Connection to a Modular IC System for Neutralization 5 30 31 32 33 Attach pump tubings x Loosen both tubing cartridges 4 from the holding clamp 9 by pressing down snap-action lever 10 and remove from mounting pin 12 on 793 Sample Prep Module (see Fig. 1). x Press contact pressure lever 5 on both tubing cartridges down as far as it will go. x Insert a length of pump tubing 33 (6.1826.010, ~1.3 mL/min) into each of the tubing cartridges as shown in Fig. 7.
2 Installation 14 6 SP Module A connection 2: Regenerant x Pull the inlet capillary 25 marked with “Regenerant“ at SP Module A connection 22 (see Fig. 6) through one of the rear panel openings 41 or 43 of the 733 IC Separation Center. x Mount a PEEK compression fitting 31 (6.2744.010) on the end of inlet capillary and screw this on to PEEK filter unit 36 (6.2821.100) (see Fig. 7).
2.3 Connection to a Modular IC System for Neutralization 8 Startup of 793 IC Sample Prep Module x Set switch 2 to "OFF". x Switch on 793 IC Sample Prep Module with mains switch 13 and set switch 3 to "ON". x Adjust the contact pressure for both tubing cartridges: press contact pressure lever 5 upwards until the solutions just start to be drawn in. Then press the contact pressure lever upwards until it clicks once more to obtain optimal contact pressure.
2 Installation 2.3.3 Settings in the «IC Net 2.1» program During configuration of the 793 IC Sample Prep Module and 753 Suppressor Module the IC Net assistant “New system wizard” requires the definition of “Event lines” for the SP Module A and suppressor, respectively. Although these lines are not installed you must enter any unused “Event line” to enable the program to proceed.
2.3 Connection to a Modular IC System for Neutralization start-neutralization.smt 2. Save the system file “neutralization.smt” with a new name “startneutralization.smt“. 3. Remove the data recorder by clicking on the recorder icon with the right mouse button an selecting 'unlink'. 4. Change this program for 'start-neutralization.
2 Installation 2.4 Connection to a Modular IC System for Cation Elimination A second application for the 793 IC Sample Prep Module is cation elimination before chromatographic separation of the sample on the column (e.g. precolumn elimination of heavy metals). In this section a system for determination of anions with chemical suppression is described. It includes the application of the sample to the SP Module A followed by filling the sample loop with this pretreated sample.
2.4 Connection to a Modular IC System for Cation Elimination 2.4.2 Connection of the SP Module A The following description implies the usage of the 2.733.0130 IC Separation Center (one injector + MSM). The 1.793.0110 SP Module A must first be inserted in the 733 IC Separation Center and connected to the 793 IC Sample Prep Module.
2 Installation 20 4 Connect SP Module A to valve A x Screw outlet capillary 29 marked with “Sample out” onto connection “1” of injection valve A using a 6.2744.010 compression fitting. 5 Attach pump tubings x Loosen both tubing cartridges 4 from the holding clamp 9 by pressing down snap-action lever 10 and remove from mounting pin 12 on 793 Sample Prep Module (see Fig. 1). x Press contact pressure lever 5 on both tubing cartridges down as far as it will go. x Insert a length of pump tubing 33 (6.
2.4 Connection to a Modular IC System for Cation Elimination 7 SP Module A connection 3: H2O x Pull the inlet capillary 28 marked with “H2O“ at SP Module A connection 22 (see Fig. 6) through one of the rear panel openings 41 or 43 of the 733 IC Separation Center. x Mount a PEEK compression fitting 31 (6.2744.010) on the end of inlet capillary 28 and screw this on to the second PEEK filter unit 36 (6.2821.100) (see Fig. 7).
2 Installation 2.4.3 Settings in the «IC Net 2.1» program During configuration of the 793 IC Sample Prep Module, the IC Net assistant “New system wizard” requires the definition of an “Event line” for the SP Module A. Although this is not installed you must enter any unused “Event line” to enable the program to proceed. Actually, the SP Module A is controlled by the 766 sample processor (see Fig. 8). The signal for switching the internal IC Suppressor Module is not provided by the 766 program.
3.1 Handling the SP Module A 3 Operation 3.1 Handling the SP Module A General The SP Module A of the 793 Sample Prep Module consists of a total of three cation exchange units which are in turn used for cation exchange, regenerated and rinsed with water. In order to record each new chromatogram under comparable conditions a freshly regenerated cation exchange unit is normally used. Switching is either carried out automatically together with valve switching or manually.
3 Operation 3.2 Manual operation The 793 IC Sample Prep Module cannot be manually operated unless the installation has been carried out properly as described in section 2 (mains connection, attaching the pump tubing, connection to the IC system). Switch instrument on/off The 793 IC Sample Prep Module is switched on and off using mains switch 13 on the rear of the instrument (see Fig.
3.3 Operation via 732 IC Detector 3.3 Operation via 732 IC Detector When operating the 793 IC Sample Prep Module together with the 732 IC Detector and the 733 IC Separation Center both pump and the SP Module A on the 793 IC Sample Prep Module can be operated and switched automatically without the program “IC Net”.
3 Operation Switch on: Switch off: >PROGRAM/edit remote XXX.X min 1✴✴✴✴✴✴✴ >PROGRAM/edit remote XXX.X min 0✴✴✴✴✴✴✴ Switch pump on/off at an event In the 732 IC Detector the key can be used to program a maximum of 4 different events (see section 4.7.3 of 732/733 Manual).
3.3 Operation via 732 IC Detector Rotate SP Module A at an event In the 732 IC Detector the key can be used to program a maximum of 4 different events (see section 4.7.3 of 732/733 Manual ). At each event the SP Module A can be brought into the next position by setting remote output lead 2 to 1 (on, active, 0 V). This lead must later be returned to the setting 0 (off, inactive, open) before the SP Module A can be switched to the next position.
4 Malfunctions - Maintenance 4 Malfunctions – Maintenance 4.1 Malfunctions and their rectification If difficulties appear with the IC system during analysis, their cause are best investigated in the order separating column o pump o eluent o IC system. An overview of possible malfunctions, their causes and remedies can be found in the instruction manuals of both 690 Ion Chromatograph (section 6.5) and 732/733 IC System (section 5.3.2).
4.1 Malfunctions and their rectification Actuator moves to next position but display lamp 7 no longer lights up Actuator is not switched to next position (display lamp 7 no longer lights up) x SP Module A dry x Rinse all 3 actuator units with solution for at least 10 min. x SP Module A temporarily in the wrong position x Switch instrument off and then on again. Switch off SP Module A remote operation with key 6 and operate key 8 . Check whether display lamp 7 lights up.
4 Malfunctions - Maintenance 4.2 Maintenance and servicing 4.2.1 Care The 793 IC Sample Prep Module requires proper care and attention. Excessive contamination of the instrument could possibly lead to malfunctions and a shorter service life of the inherently rugged mechanical and electronic parts. Spilled chemicals and solvents should be wiped up immediately. It is especially important to protect the plug connections at the rear of the instrument (particular the mains plug) against contamination.
4.2 Maintenance and servicing 4.2.4 Exchanging the pump tubing Pump tubings are consumable material with a limited lifetime and should be exchanged at regular intervals (approx. every 4 weeks under continuous use). The working life of pump tubing depends to a considerable extent on the contact pressure. This is why the contact pressure must be correctly set as described in section 2.3.2 or 2.4.2 respectively.
4 Malfunctions - Maintenance 4.2.5 Regeneration of the cation exchange material If the actuator units are exposed to certain heavy metals (e.g. iron) or organic contaminants for longer periods of time then these can no longer be removed completely by the regeneration solution normally used (20 mmol/L HClO4). This diminishes the capacity of the actuator units.
4.2 Maintenance and servicing 4.2.6 Cleaning the SP Module A Cleaning the SP Module A may be necessary in the following cases (see also section 4.1): x Blockage of the SP Module A which cannot be remedied (the SP Module A can no longer deliver solutions) x Obstruction of the actuator which cannot be remedied (the actuator can no longer be switched to next position) To clean the connection piece and the rotor proceed as follows (see Fig.
4 Malfunctions - Maintenance 37 38 39 40 Fig. 9: Assembling the actuator 39 37 Screw nut 38 Connection piece (6.2835.010) 40 with input and output leads Actuator rotor (6.2835.000) Actuator holder 6 Clean connection piece x Clean the sealing surface of connection piece 38 with the aid of a lint-free cloth and ethanol.
4.2 Maintenance and servicing 4.2.7 Exchanging the cation exchange material Exchanging the cation exchange material in the SP Module A of the 793 IC Sample Prep Module may be necessary in the following cases (see also section 4.1): x Loss of cation exchange capacity which cannot be remedied x Blockage of the SP Module A which cannot be remedied Both the 6.2835.000 actuator rotor as well as the 6.2835.010 connection piece with the input and output leads can be exchanged.
4 Malfunctions - Maintenance 7 Connect and condition the SP Module A x Reconnect the actuator to the IC system (see section 2.3.2 and section 2.4.2). x Before switching the actuator to the next position for the first time rinse all 3 actuator units with solution for 5 min. 8 Check braking range x Check the braking range of the SP Module A according to section 4.2.8 and readjust if necessary. 4.2.
4.2 Maintenance and servicing 3 Adjust braking range x Rotate adjustment screw 18 a little in each direction according to Fig. 10 so that the rotor position lies in the acceptable range. x Operate key 8 . x Check the position of the rotor in all of the 3 possible positions according to Fig. 10. x Repeat the adjustment procedure until the position of the rotor lies in the acceptable range in all of the 3 possible positions. Wait at least 10 s each time before operating key 8 .
5 Appendix 5 Appendix 5.1 Technical data Pump Pump type 2-channel peristaltic pump with rotational speed of 20/min (50 Hz) 24/min (60 Hz) Pump capacity typ. ~1.3 mL/min with 6.1826.010 pump tubing Pressure max. 4 bar (0.
5.1 Technical data Mains connection Mains voltage 115 V: 100...120 V r 10 % 230 V: 220...240 V r 10 % Switchable with voltage selector in fuse holder (see section 2.2.1) Mains frequency 50...60 Hz Power consumption 50 VA Fuse 5 mm Ø, 20 mm length 100}120 V: 0.5 A (slow-blow) 220}240 V: 0.
5 Appendix 5.2 Standard equipment Subject to changes! All dimensions are given in mm. The 2.793.0010 Sample Prep Module includes the following parts: Quant. Order No. Description 1 1.793.0010 IC Sample Prep Module 1 1.793.0110 SP Module A 1 6.1803.020 PTFE capillary tubing Length = 5 m 2 6.1826.010 Pump tubing made of PVC (Tygon£); with 2 firmly attached white-white stoppers; i.d. = 1.02 mm, e.d. = 2.72 mm 1 6.2122.0X0 0.97 1.
5.3 Optional accessories 2 6.2744.110 PEEK coupling Connection between 6.2821.100 PEEK filter unit and 6.1826.010 pump tubing. 33 2 6.2821.100 PEEK Filter Unit 2 Pm To avoid contamination due to abrasive particles of piston seals 16 Spare part: 6.2821.110 Filter 25.5 1 5.3 8.793.1003 Instructions for use (English) for 793 IC Sample Prep Module Optional accessories Order No. Description 6.2125.090 Cable RS232 extension cable (1:1); with spezial shielding 6.2125.
5 Appendix 5.4 Warranty and conformity 5.4.1 Warranty The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error which occur within 12 months from the day of delivery. In such cases the defects will be rectified in our workshops free of charge. Transport costs are to be paid by the customer. For day and night operation the warranty is limited to 6 months. Glass breakage in the case of electrodes or other parts is not covered by the warranty.
5.4 Warranty and conformity 5.4.2 EU Declaration of conformity EU Declaration of Conformity The METROHM AG company, Herisau, Switzerland hereby certifies, that the instrument: 793 IC Sample Prep Module meets the requirements of EC Directives 89/336/EWG and 73/23/EWG.
5 Appendix 5.4.3 Certificate of conformity and system validation Certificate of Conformity and System Validation This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company. Name of commodity: 793 Sample Prep Module Name of manufacturer: Metrohm Ltd.
5.5 Index 5.5 Index A Actuator......................... see SP Module A Actuator holder 40 Assembling ....................................... 35 Figure ................................................ 34 Actuator rotor 39 Check position ............................. 36,37 Cleaning....................................... 33,35 Exchange .......................................... 35 Figure ................................................ 34 Insertion .......................................
5 Appendix Optional accessories............................ 41 Organic contamination......................... 32 Organization ............................................4 Outlet capillary for acid 26 Connection .................................. 14,21 Figure................................................ 12 Outlet capillary for H2O 27 Connection .................................. 15,21 Figure................................................ 12 Outlet capillary for sample 29 Connection .....................