User guide
2 Installation
788 IC Filtration Sample Processor
22
x
Pull the PEEK capillary
20
2020
20
through the opening
22
2222
22
or
28
2828
28
of
the 733 IC Separation Center and screw onto connection "1"
of injection valve
66
6666
66
using a 6.2744.010 PEEK compression
fitting
6
66
6
.
x
Retighten rotary nipple on the interior side of connection
22
2222
22
or
28
2828
28
to fix the capillary
20
2020
20
.
7 Tubing connection injection valve – waste
x
Insert 6.2744.020 coupling (from 733 accessories) into
connection
21
2121
21
or
27
2727
27
of the 733 IC Separation Center.
x
Screw PTFE suction tubing
84
8484
84
onto the 6.2744.020 coupling
attached to connection
21
2121
21
or
27
2727
27
and lead it into the waste
container.
In the case of the 733.0020 IC Separation Center with two injection
valves, it is possible to fill both sample loops from the same 788 IC
Filtration Sample Processor. For this, connection "1" of valve A (outlet
of the sample loop) must be connected to connection "2" of valve B
(inlet of the sample loop) using a 6.1803.040 PEEK capillary (15 cm).
2.3.10 Tubing connections 788 – 754 Dialysis Unit
If the 788 IC Filtration Sample Processor is used for an IC system with
sample dialysis (see
section 2.4.7 or section 2.5.8
), the peristaltic pump
at the 754 Dialysis Unit is used for sample conveying instead of the
pump at the 788 IC Filtration Sample Processor. Additionally, the ultra-
filtration cell is bridged. The tubing connections between the 754
Dialysis Unit and the 733 IC Separation Center have to be made as
shown in
Fig. 8
of the
754 Instruction for Use
. The only change concerns
the inlet of the sample, which is installed as follows:
Install a PEEK compression fitting 6.2744.010 on the inlet end of the
PTFE tubing
26
2626
26
(6.1803.030, see
Fig. 8 of the 754 Instruction for Use
)
connected to the 754 Dialysis Unit and screw it on the PEEK
compression fitting
5
55
5
(see
Fig. 6
).
2.3.11 Fixing tubing and cables
In order to fix tubing or cables in the
guide chain
9
99
9
any chain link may be
opened with a screw driver or another
appropriate tool..
Fig. 12
: Opening chain links










