User guide

2.3 Attaching the accessories
788 IC Filtration Sample Processor
19
2.3.9 Tubing connections
The tubing connections on the 788 IC Filtration Sample Processor and
the connection to the injection valve of the 733 IC Separation Center
must be made as follows:
Pump tubings are consumable material with a lifetime which depends
on the contact pressure (see chapter 6.3.3). This is why the tubing
cartridges should be raised completely by loosening snap-action lever
16
1616
16
on the right-hand side if the pump is to remain switched off for a
considerable length of time (the set contact pressure remains
unchanged).
The 6.1826.0X0 pump tubing is made of PVC and must not be used
for rinsing with solutions which contain acetone. In such cases, rinse
with different pump tubing or a different pump.
8/21
8/218/21
8/21
19
1919
19 10/11
10/1110/11
10/11 43
4343
43 14
1414
14 18
1818
18 43
4343
43 19
1919
19
24/20
24/2024/20
24/20
16
1616
16 6
66
66
66
6
Abb. 10
: Installing the pump tubing
Line 1: Conveying the sample
to the ultra-filtration cell
Line 2: Conveying the filtrate
to the injection loop
8
88
8
PTFE capillary (6.1803.070)
21
2121
21
PTFE capillary (6.1803.050)
6
66
6
PEEK compression fitting
(6.2744.010)
6
66
6
PEEK compression fitting (6.2744.010)
19
1919
19
PEEK coupling (6.2744.030)
19
1919
19
PEEK coupling (6.2744.030)
10
1010
10
Pump tubing (6.1826.070)
11
1111
11
Pump tubing 6.1826.030
43
4343
43
Stopper (yellow/yellow)
43
4343
43
Stopper (orange/yellow)
14
1414
14
Tubing cartridge (6.2755.000)
14
1414
14
Tubing cartridge (6.2755.000)
18
1818
18
Contact pressure lever
18
1818
18
Contact pressure lever
16
1616
16
Snap-action lever
16
1616
16
Snap-action lever
24
2424
24
PTFE capillary (6.1803.060)
20
2020
20
PEEK capillary (6.1831.060)