User guide
2 Installation
788 IC Filtration Sample Processor
12
2.3 Attaching the accessories
For attaching the accessories at the 788 IC Filtration Sample
Processor, proceed in the order described below.
2.3.1 Connecting the swing head
Plug in the branch plug of the connection cable
29
29 29
29
permanently
attached to the swing head into the remote connection socket at the
788 IC Filtration Sample Processor and screw it onto this connection
using a screw-driver (see
Fig. 1
).
2.3.2 Connecting the keyboard
1
2
Connecting the keyboard
Connect the 6.2142.010 keyboard
to the keyboard connection
39
3939
39
"Keyboard". For disconnection,
press the plug together slightly on
both sides.
Switch on instrument
Switch on the 788 IC Filtration
Sample Processor with mains
switch
34
3434
34
. The keyboard display
lights up. The instrument is
initialized and the lift is raised
completely.
Fig. 4
: Keyboard connection
2.3.3 Installing the plug cover
To prevent any contamination of the mains and remote connection by
spilled solvents or chemicals, the 6.2752.010 plug cover must always
be installed when operating the 788 IC Filtration Sample Processor!
Install the plug cover
31
3131
31
in the corresponding guide groove above
mains connection plug
33
3333
33
and remote connection
30
3030
30
(see
Fig. 1
and
Fig. 2
).