Do not connect the dialysis cell with anything other than the
6.2744.000 PVDF compression fittings supplied. If 6.2744.010
PEEK compression fittings are used, stress cracks may appear in the
ultra-filtration cell !
4 Close ultra-filtration cell
x
Insert sealing ring in the appropriate recess in top part
43
4343
43
.
x
Place top part on bottom part
45
4545
45
so that both parts lie flush.
x
Using the 6.2621.070 Allen key, screw the 5 screws
46
46 46
46
right
in and tighten up.
5 Attach ultra-filtration cell
x
Insert the assembled ultra-filtration cell in the cell holder
26
26 26
26
according to Fig. 1 or Fig. 11. The heads of the five screws
46
46 46
46
fit to the small holes of the cell holder.
x
Connect all three tubings according to step 3 to 5 in
section
2.3.9
to the corresponding inlet and outlet openings of the ul-
tra-filtration cell by using PVDF compression fittings
22
2222
22
.
6 Rinse ultra-filtration cell
x
Every time the filtration membrane is changed, the remaining
air in the cell and tubings has to be removed. This can be
done by rinsing with ultrapure water.
x
A completely installed 788 IC Filtration Sample Processor with
ultra-filtration cell is required for processing the following
steps. The rinsing solution (e.g. ultrapure water) is stored in a
300 mL PE bottle on the special beaker position 1. The 788 IC
Filtration Sample Processor is online and in normal state. The
pump is switched of (display: '
PUMP–'
) and the 788 is ready
for use (display: '
ready
').
x
Turn the sample rack to position the beaker with the rinsing
solution under the needle by pressing <MOVE> followed by
choosing 's
pec.1
' with <SELECT>.
x
Lower the lift into rinse position with <LIFT> and <SELECT>
'
rinse
'
x
Turn on the pump by pressing <PUMP>. All connected tub-
ings and the ultra-filtration cell will be rinsed.
x
After adequate time (e.g. 5 min.), switch off the Pump
(<PUMP>) and raise the lift to its rest position by pressing
<HOME>.