User Manual
3.2 Remote control commands
768 KF Oven
21
│& Root
│ Info Information
│ │ .Report Sending of formatted reports $G 3.2.2.17
│ │ │ .Select Selection of the report configuration,
│ │ │ parameters,
│ │ │ result 3.2.2.18
│ │ .Results Results of the determination
│ │ │ .PurgeTime Purge time read only 3.2.2.19
│ │ │ .CondTime Conditioning time read only 3.2.2.19
│ │ │ .SmplHeatTime Sample heating time read only 3.2.2.19
│ │ │ .LowTemp Lowest temp. during heating read only 3.2.2.19
│ │ │ .HighTemp Highest temp. during heating read only 3.2.2.19
│ │ │ .GasFlow Mean gas flow read only 3.2.2.19
│ │ │ .LowFlow Lowest flow during heating read only 3.2.2.19
│ │ │ .HighFlow Highest flow during heating read only 3.2.2.19
│ │
│ │ .ActualInfo Current information
│ │ │ .Inputs I/O lines, inputs
│ │ │ │ .Status Status of the lines read only 3.2.2.20
│ │ │ │ .Change Change in the status of the lines read only 3.2.2.20
│ │ │ │ .Clear Clears the change byte read only 3.2.2.20
│ │ │ .Outputs I/O lines, outputs
│ │ │ │ .Status Status of the lines read only 3.2.2.20
│ │ │ │ .Change Changing the status of the lines read only 3.2.2.20
│ │ │ │ .Clear Clears the change byte read only 3.2.2.20
│ │ │ .Meas Measured values
│ │ │ │ .CyclNo Cycle number read only 3.2.2.21
│ │ │ │ .SampleTemp Sample temperature read only 3.2.2.21
│ │ │ │ .OvenTemp Oven temperature read only 3.2.2.21
│ │ │ │ .GasFlow Gas flow read only 3.2.2.21
│ │ │ .Status Status of the assemblies
│ │ │ │ .BoatPos Position of the sample boat read only 3.2.2.22
│ │ │ │ .Valve Valve position read only 3.2.2.22
│ │ │ │ .Pump Pump read only 3.2.2.22
│ │ │ │ .Heating Heating power level read only 3.2.2.22
│ │ │ .Display Display
│ │ │ │ .L1 1st line 24 ASCII 3.2.2.23
│ │ │ │ .L2 2nd line 24 ASCII 3.2.2.23
│ │
│ │ .Assembly Assembly
│ │ │ .CycleTime Cycle time read only 3.2.2.24
│
│
│ Assembly Basic components of the assembly
│ │ .Prep Perform preparations $G, $S 3.2.2.25
│ │ .Heat Heating on $G 3.2.2.26
│ │ │ .Value Heating power level 0...50 3.2.2.26
│ │ .Valve Set valve position $G 3.2.2.27
│ │ │ .Pos Preselect position purge, transfer 3.2.2.27
│ │ .Boat Move sample boat $G, $S 3.2.2.28
│ │ │ .Rate Moving rate 0.1...10 3.2.2.28
│ │ │ .Pos Go to position 0...130.0 3.2.2.28
│ │ │ .SetPos Set stop points
│ │ │ │ .InPos Inner stop 0...130.0 3.2.2.28
│ │ │ │ .OutPos Outer stop 0...130.0v 3.2.2.28
│ │ .Pump Switch pump on/off $G, $S 3.2.2.29
│ │ .Outputs I/O lines, outputs
│ │ │ .SetLines Set lines $G 3.2.2.30
│ │ │ │ .L1 Line 1 active, inactive
│ │ │ │ : pulse, OFF 3.2.2.30
│ │ │ │ .L8 All lines as L1
│ │ │ .ResetLines Set all lines inactive $G 3.2.2.30