User Manual
3.2 Remote control commands3.2 Remote control commands
768 KF Oven
20
3.2 Remote control commands
3.2.1 Overview
│& Root
│ Mode Mode $G, $S 3.2.2.1
│ │ .Temp Sample temperature 50...300 3.2.2.2
│ │ .Gas Gas flow
│ │ │ .UnitFlow Unit for display mL/min, L/h 3.2.2.3
│ │ │ .MinFlow Minimum gas flow 0...999 3.2.2.3
│ │ │ .Type
│ │ │ │ .Select Selection of the gas type air, N2, other 3.2.2.4
│ │ │ │ .OtherFac Factor for "other" gas 0.001...9.999 3.2.2.4
│ │ .PurgeTime Purge time 0...99999 3.2.2.5
│ │ .CondTime Conditioning time 0...99999 3.2.2.5
│
│
│ Config Configuration
│ │ .OvenSet Oven settings
│ │ │ .AutoPrep Automatic preparation ON, OFF 3.2.2.6
│ │ │ .ValveControl Valve control ON, OFF 3.2.2.7
│ │ │ .StartCond Start allowed only if cond.ok ON, OFF 3.2.2.7
│ │ │ .TempLimit Starting temperature range 1...100 3.2.2.7
│ │ │ .TempCorr Correction of sample temperature 0...±99.9 3.2.2.7
│ │ │ .CharSet Selection of the character set Epson, Seiko,
│ │ │ Citizen, HP, IBM 3.2.2.8
│ │ │ .Report Report output at end ON, OFF 3.2.2.9
│ │
│ │ .Aux Setting of various auxiliary functions
│ │ │ .Language Selection of the dialog language english, deutsch,
│ │ │ francais,
│ │ │ espanol 3.2.2.10
│ │ │ .RunNo Current run number 0...9999 3.2.2.11
│ │ │
│
│ │ .AutoStart Internal automatic start 1...9999,OFF 3.2.2.12
│ │ │ .StartDelay Start delay 0...9999 3.2.2.12
│ │ │ .Beeper Beep 1...9, OFF 3.2.2.13
│ │ │ .DevName Device name 8 ASCIII 3.2.2.13
│ │ │ .Prog Program version read only 3.2.2.14
│ │
│ │ .RSSet Settings for RS232 $G 3.2.2.15
│ │ │ .Baud Baud rate (300...9600) 3.2.2.16
│ │ │ .DataBit Data bits 7, 8 3.2.2.16
│ │ │ .StopBit Stop bits 1, 2 3.2.2.16
│ │ │ .Parity Parity even, odd,
│ │ │ none 3.2.2.16
│ │ │ .Handsh Handshake HWs, HWf,
│ │ │ SWchar, SWline,
│ │ │ none 3.2.2.16