User Manual

3.1 General rules
768 KF Oven
19
RS send errors:
E40 DSR=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
E41 DCD=ON. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
E42 CTS=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
E43 The transmission has been interrupted with XOFF for at least 3
s.
Exit: Send XON or <QUIT>.
E44 The RS parameters are no longer the same for both devices.
Reset.
E45 The receive buffer contains an incomplete command (L
F
missing). Sending from the instrument is therefore blocked.
Exit: Send L
F
or <QUIT>.
Device-specific errors:
E26 Manual stop.
Exit: The error message disappears on the next start.
E135 Check temperature sensor of sample temperature.
Exit: Rectify fault.
E154 Sample temperature not OK.
Exit: The error message disappears when the fault is rectified
or &M $S.
E163 Gas flow too low.
Exit: The error message disappears when the fault is rectified
or &M $S.
E164 The attached titrator is not conditioned.
Exit: The error message disappears when the fault is rectified
or &M $S.
E165 The oven temperature is higher than 360 °C.
Exit: The error message disappears when the fault is rectified.
E168 The temperature sensor for the oven temperature is not OK.
Exit: Rectify fault.
E169 The gas flow sensor is not OK (flow >500 mL/min).
Exit: Rectify fault. Continuation is possible only when the
minimum gas flow is set to 0 mL/min.