User Manual
3.1 General rules3.1 General rules
768 KF Oven
18
3.1.3 Status and error messages
Detailed status conditions
Status conditions of the global $G:
$G.Mode .Inac Waiting during start delay
.PurgeTime Waiting during purge time
.CondTime Waiting during cond.time
.HeatSmpl Heating the sample
.Terminate Carrying out the terminating steps
$G.Assembly .Prep.Wait Heating the oven to the set temperature
.Boat The sample boat has been manually moved
Status conditions of the global $R:
$R.Mode .Ready Inactive: Ready to start an automatic determination
$R.Assembly.Ready An Assembly step has been carried out. Ready to carry out
another Assembly step (&Mode $G will trigger error message
E31).
Status conditions of the global $S:
The instrument gives the status from which it has been stopped. The detailed status information is
therefore identical as for the global status $G.
Error messages
RS receive errors:
E36 Parity error
Exit: <QUIT> and set parity at both devices the same.
E37 Stop Bit
Exit: <QUIT> and set stop bit at both devices the same.
E38 Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
E39 The internal receive buffer is full (>82 characters).
Exit: <QUIT>.