Manual

768 KF Oven 3
Rear
10 Tubing connection
with 6.1805.040 Tubing.
11 Contrast adjustment for the display
12 Dust filter
13 Connection for titrators
14 Connection for temperature sensor
plug the grey connector of the cable into the red socket of the
KF Oven!
15 Connection for outlet heating
16 RS232 interface
for connection of a printer or computer.
17 Model plate
with manufacturing, serial and instrument number.
Display of the set mains voltage, the mains frequency and the
power consumption. Before switching on for the first time,
check that the set mains voltage matches the voltage of your
power supply. If this is not the case, disconnect mains cable
and change over voltage, see page 43.
18 Power supply unit
with mains switch and mains connection.
With power supplies in which powerful HF disturbances are
superimposed on the mains voltage, the KF Oven should be
operated with an additional mains filter, e.g. METROHM model
615.