Manual
5.4 Connection of the outlet heating
768 KF-Ofen
50
5.4 Connection of the outlet heating
If the 6.1830.000 Heatable Outlet Tubing is used, appreciably shorter determination times
are achieved.
The heatable outlet tubing should be used in all cases where problems arise with con-
densation water in the normal outlet tubing.
Screw tubing to the outlet connection and connect cable of the outlet heating to the KF
Oven:
707
A B C
D
Kabel der Auslassheizung
The heatable outlet tubing is inserted in the titration vessel as follows:
Septum Stopper
6.2730.020
6.1830.000 Outlet
Tubing
6.1830.000 Outlet
Tubing
6.1830.000 Outlet
Tubing
E.301.0032
O-ring
E.301.0032
O-ring
Stopper
6.1446.170
incl.
O-ring
E.301.0032
Screw Nipple
6.2730.010
• If you use the outlet heating together with the 756 KF Coulometer in a lateral aperture
of the titration vessel, use the A.254.0102 sealing ring instead of A.254.0104 (see sec-
tion 5.2).
768
756 KF Coulometer
or
737 KF Coulometer,
cell without diaphragm
737 KF Coulometer,
cell with diaphragm
Volumetric KF titration
vessel
Cable of outlet heating