Manual
5.1 Setting up the Oven
768 KF-Ofen
45
5.1 Setting up the Oven
Unpack Oven and remove all protective coverings.
1. Insert 6.2407.020 Insert Tube in the heating chamber of the Oven and fasten it to the
Oven with the clamp (red screw). The glass protrusion on the insert tube is positioned in
the slot of the clamp.
2. Screw 6.2104.080 Cable onto temperature sensor of the insert tube and connect to the
Pt1000 sockets of the Oven.
Important: The grey connector of the cable is plugged into the red (upper) socket of the
Oven!
3. Place a 6.1448.040 Septum in the 6.2701.060 Screw Cap and screw it to the thread of
the insert tube.
Close the insert tube with the 6.2750.010 Cover.
4. Fill the 6.1608.050 Drying Flasks with molecular sieves.
Screw the 6.1821.040 Inlet Tubes onto the 6.1602.140 Screw Cover.
Screw covers onto drying flasks.
5. Place drying flasks in the holder of the Oven.
Screw 6.1805.080 Tubing to the inlet tube of the left drying flask, see drawing on holder.
Screw other end of the tubing "To drying flask" at the Oven.
6. Connect the left drying flask to the right drying flask using 6.1805.180 Tubing: On the
right flask, screw the tubing onto the inlet tube.
6.1805.080
6.1805.180
6.1805.180
6.1805.070
6.1808.130
(6.1808.140)