Manual

4.1 Troubleshooting
38 768 KF Oven
Problem Remedial action
Titration times differ
widely.
Grind sample before weighing and spread over
sample boat as evenly as possible.
Position the Oven at a location where it is protected
against draughts and direct sunlight.
The titrator does not
switch off.
Grind sample before weighing and spread over
sample boat as evenly as possible.
The sample decomposes and evolves oxidizable
substances: Lower temperature, possibly use N2 as
carrier gas.
Gas does not reach
titration vessel.
Tube blocked by molecular sieve?
Inlet of a drying flask blocked by molecular sieve?
Switch valve to "TRANSFER" and possibly increase
gas flow.
Are the cover of the insert tube and the sample inlet
at the top closed properly?
Are all tubing connections tight?
No gas flow although the
pump is adjusted or
another gas is
connected.
Tube blocked by molecular sieve?
Inlet of a drying flask blocked by molecular sieve?
Firmly tighten all tubing connections.
Has the tubing been connected at the rear of the
oven?
Error message
"conditioning not ok"
although the attached
titrator is conditioned.
Is the connecting cable to the titrator plugged in
properly?
On interconnection with Titrinos, check that the
cable ends have not been mixed up!