3.2 Remote control commands3.2 Remote control commands
768 KF Oven
30
3.2.2.41 Setup.RamInit $G
Initialises the device as in the diagnostic test, see page 42: All parameters are set to the
default value and error messages are cleared.
3.2.2.42 Setup.InstrNo $G
Setup.InstrNo.Value up to 8 ASCII
Device identification outputted in the report. The serial number and manufacturing number
are entered here in the factory as an unequivocal identification.
3.2.2.43 Setup.Save $G
All entries made via RS232 must be saved with &Setup.Save $G before the instrument is
switched off to avoid losses.