Manual
3.2 Remote control commands
768 KF Oven
21
³& Root
³ Info Information
³ ³ .Report Sending of formatted reports $G 3.2.2.17
³ ³ ³ .Select Selection of the report configuration,
³ ³ ³ parameters,
³ ³ ³ result 3.2.2.18
³ ³ .Results Results of the determination
³ ³ ³ .PurgeTime Purge time read only 3.2.2.19
³ ³ ³ .CondTime Conditioning time read only 3.2.2.19
³ ³ ³ .SmplHeatTime Sample heating time read only 3.2.2.19
³ ³ ³ .LowTemp Lowest temp. during heating read only 3.2.2.19
³ ³ ³ .HighTemp Highest temp. during heating read only 3.2.2.19
³ ³ ³ .GasFlow Mean gas flow read only 3.2.2.19
³ ³ ³ .LowFlow Lowest flow during heating read only 3.2.2.19
³ ³ ³ .HighFlow Highest flow during heating read only 3.2.2.19
³ ³
³ ³ .ActualInfo Current information
³ ³ ³ .Inputs I/O lines, inputs
³ ³ ³ ³ .Status Status of the lines read only 3.2.2.20
³ ³ ³ ³ .Change Change in the status of the lines read only 3.2.2.20
³ ³ ³ ³ .Clear Clears the change byte read only 3.2.2.20
³ ³ ³ .Outputs I/O lines, outputs
³ ³ ³ ³ .Status Status of the lines read only 3.2.2.20
³ ³ ³ ³ .Change Changing the status of the lines read only 3.2.2.20
³ ³ ³ ³ .Clear Clears the change byte read only 3.2.2.20
³ ³ ³ .Meas Measured values
³ ³ ³ ³ .CyclNo Cycle number read only 3.2.2.21
³ ³ ³ ³ .SampleTemp Sample temperature read only 3.2.2.21
³ ³ ³ ³ .OvenTemp Oven temperature read only 3.2.2.21
³ ³ ³ ³ .GasFlow Gas flow read only 3.2.2.21
³ ³ ³ .Status Status of the assemblies
³ ³ ³ ³ .BoatPos Position of the sample boat read only 3.2.2.22
³ ³ ³ ³ .Valve Valve position read only 3.2.2.22
³ ³ ³ ³ .Pump Pump read only 3.2.2.22
³ ³ ³ ³ .Heating Heating power level read only 3.2.2.22
³ ³ ³ .Display Display
³ ³ ³ ³ .L1 1st line 24 ASCII 3.2.2.23
³ ³ ³ ³ .L2 2nd line 24 ASCII 3.2.2.23
³ ³
³ ³ .Assembly Assembly
³ ³ ³ .CycleTime Cycle time read only 3.2.2.24
³
³
³ Assembly Basic components of the assembly
³ ³ .Prep Perform preparations $G, $S 3.2.2.25
³ ³ .Heat Heating on $G 3.2.2.26
³ ³ ³ .Value Heating power level 0...50 3.2.2.26
³ ³ .Valve Set valve position $G 3.2.2.27
³ ³ ³ .Pos Preselect position purge, transfer 3.2.2.27
³ ³ .Boat Move sample boat $G, $S 3.2.2.28
³ ³ ³ .Rate Moving rate 0.1...10 3.2.2.28
³ ³ ³ .Pos Go to position 0...130.0 3.2.2.28
³ ³ ³ .SetPos Set stop points
³ ³ ³ ³ .InPos Inner stop 0...130.0 3.2.2.28
³ ³ ³ ³ .OutPos Outer stop 0...130.0v 3.2.2.28
³ ³ .Pump Switch pump on/off $G, $S 3.2.2.29
³ ³ .Outputs I/O lines, outputs
³ ³ ³ .SetLines Set lines $G 3.2.2.30
³ ³ ³ ³ .L1 Line 1 active, inactive
³ ³ ³ ³ : pulse, OFF 3.2.2.30
³ ³ ³ ³ .L8 All lines as L1
³ ³ ³ .ResetLines Set all lines inactive $G 3.2.2.30