Manual
3.2 Remote control commands3.2 Remote control commands
768 KF Oven
20
3.2 Remote control commands
3.2.1 Overview
³& Root
³ Mode Mode $G, $S 3.2.2.1
³ ³ .Temp Sample temperature 50...300 3.2.2.2
³ ³ .Gas Gas flow
³ ³ ³ .UnitFlow Unit for display mL/min, L/h 3.2.2.3
³ ³ ³ .MinFlow Minimum gas flow 0...999 3.2.2.3
³ ³ ³ .Type
³ ³ ³ ³ .Select Selection of the gas type air, N2, other 3.2.2.4
³ ³ ³ ³ .OtherFac Factor for "other" gas 0.001...9.999 3.2.2.4
³ ³ .PurgeTime Purge time 0...99999 3.2.2.5
³ ³ .CondTime Conditioning time 0...99999 3.2.2.5
³
³
³ Config Configuration
³ ³ .OvenSet Oven settings
³ ³ ³ .AutoPrep Automatic preparation ON, OFF 3.2.2.6
³ ³ ³ .ValveControl Valve control ON, OFF 3.2.2.7
³ ³ ³ .StartCond Start allowed only if cond.ok ON, OFF 3.2.2.7
³ ³ ³ .TempLimit Starting temperature range 1...100 3.2.2.7
³ ³ ³ .TempCorr Correction of sample temperature 0...±99.9 3.2.2.7
³ ³ ³ .CharSet Selection of the character set Epson, Seiko,
³ ³ ³ Citizen, HP, IBM 3.2.2.8
³ ³ ³ .Report Report output at end ON, OFF 3.2.2.9
³ ³
³ ³ .Aux Setting of various auxiliary functions
³ ³ ³ .Language Selection of the dialog language english, deutsch,
³ ³ ³ francais,
³ ³ ³ espanol 3.2.2.10
³ ³ ³ .RunNo Current run number 0...9999 3.2.2.11
³ ³ ³
³ ³ ³ .AutoStart Internal automatic start 1...9999,OFF 3.2.2.12
³ ³ ³ .StartDelay Start delay 0...9999 3.2.2.12
³ ³ ³ .Beeper Beep 1...9, OFF 3.2.2.13
³ ³ ³ .DevName Device name 8 ASCIII 3.2.2.13
³ ³ ³ .Prog Program version read only 3.2.2.14
³ ³
³ ³ .RSSet Settings for RS232 $G 3.2.2.15
³ ³ ³ .Baud Baud rate (300...9600) 3.2.2.16
³ ³ ³ .DataBit Data bits 7, 8 3.2.2.16
³ ³ ³ .StopBit Stop bits 1, 2 3.2.2.16
³ ³ ³ .Parity Parity even, odd,
³ ³ ³ none 3.2.2.16
³ ³ ³ .Handsh Handshake HWs, HWf,
³ ³ ³ SWchar, SWline,
³ ³ ³ none 3.2.2.16