Manual

2.3 Settings on KF Oven
768 KF Oven
12
2.3.4 Special settings
>lock
<configuration> OFF
<parameters> OFF
<heater> OFF
<pump> OFF
<valve> OFF
<boat> OFF
Locking keys
Key <CONFIG> (ON, OFF)
Key <PARAM> (ON, OFF)
Key <HEATER> (ON, OFF)
Key <PUMP> (ON, OFF)
Key <VALVE> (ON, OFF)
Key <BOAT IN> and <BOAT OUT> (ON, OFF)
>T controller
init.heat.factor 100 %
add.heat.factor 100 %
Factors for T controller of Oven
The controller is optimized, therefore it normally isn't nec-
essary to change the following factors.
Initial heating factor (0...200 %)
Operative after the sample has been moved into the Oven.
Set a lower value if the temperature is below the nominal
temperature, and much above afterwards.
Additional heating factor (0...200 %)
After the initial heating, the heating is controlled by meas-
urements.
Set a greater value if the temperature is not reached dur-
ing too long a time. In this case, a higher "init.heat.factor"
may also help.
Switch the Oven OFF. Press key <CONFIG>, keep key
pressed and switch the Oven ON again.
CONFIG
+Switching ON