Manual

2.3 Settings on KF Oven
768 KF Oven
10
>auxiliaries
dialog: english
run number 0
auto start OFF
start delay 0 s
beeper 1
device label
program 768.0012
Inquiry group: Auxiliaries
Selection of the dialog language (english, deutsch, fran-
cais, espaƱol)
Current run number (0...9999)
The run number is set to 0 when the oven is switched on.
Automatic, internal apparatus start (1...9999, OFF)
Number of automatic starts.
Start delay (0...9999 s)
Delay time following the start before the automatic deter-
mination begins. The wait time can be aborted with
<QUIT>.
Audio signal (1...9, OFF)
Number of beeps when the oven is ready.
Device label for the individual identification of devices in
the device network (up to 8 ASCII characters)
Display of the program version
>RS232 settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
RS control: ON
Inquiry group: RS settings
Baud rate (300, 600, 1200, 2400, 4800, 9600)
Data bits (7, 8)
Stop bits (1, 2)
Parity (even, odd, none)
Handshake (HWs, HWf, SWline, SWchar, none)
see also page 31 ff.
Control via RS232 interface (OFF, ON)
"OFF" means the receipt of commands and data via the
RS232 interface is blocked. Data output is possible.