2.3 Settings on KF Oven
768 KF Oven
9
2.3.2 <CONFIG> key
>oven settings
auto preparation: OFF
valve control: ON
start if cond.ok: OFF
start temp.range 5 °C
temp.correcture 0 °C
send to: IBM
report: OFF
Inquiry group: General oven settings
Automatic preparation after switching on (ON, OFF)
"ON " means: Automatic heating to the sample tempera-
ture.
Valve control (ON, OFF)
"ON" means automatic switching of the valve from
"PURGE" to "TRANSFER" at the end of the determination.
With "OFF", the valve remains on "TRANSFER".
Start possible only when the attached titrator is condi-
tioned (ON, OFF)
"ON" means: The sample boat is not moved to the hot part
of the oven until the attached titrator is conditioned.
If no titrator is attached, "OFF" must be set.
Temperature range for starting (1...100 °C)
Starting the determination is only possible if the current
temperature is in the range of ±X °C from the sample
temperature.
Temperature correction (0...±99.9 °C)
Correction for sample temperature, see page 55.
Selection of the printer type/character set (Epson, Seiko,
Citizen, HP, IBM)
"Epson" for Epson Mode
"Seiko", e.g. for DPU-414
"Citizen", e.g. for i560 RS
"HP" for HP mode
"IBM" for all printers with IBM character set table 437, as
well as for data transmission to a computer or a data sys-
tem.
Report output at the end of the automatic determination
(OFF, ON)
Report example:
'fr
768 KF Oven 0D1/108 768.0010
run number 1
purge time 10 s
cond.time 5 s
smpl heating time 587 s
sample temp. 150 °C Set temperature
lowest temp. 147 °C During sample heating
highest temp. 150 °C During sample heating
gas type: air
gas flow 87 mL/min Mean gas flow
=============
The inquiries are divided into groups. The titles of the in-
quiry groups are marked with the ">" character. Entry into
the inquiry group with <ENTER>.
CONFIG