User Manual

5 Interfaces
766 IC Sample Processor
92
CTL commands
Parameter Bit pattern Function Signal
INIT 00000000000000 initializes the remote interface static
INIT 732 ***0000*000**0 initializes the remote lines at 732/1 and 732/2 static
PROG R/S 1 ***000*******1 starts/stops time program at 732/1 pulse (200 ms)
PROG R/S 2 ******0*100*** starts/stops time program at 732/2 pulse (200 ms)
PUMP R/S 1 ***001*******0 starts/stops 709/1 IC Pump pulse (200 ms)
FILL A 1 ***010*******0 switches valve A at 733/1 to "Fill" pulse (200 ms)
INJECT A 1 ***100*******0 switches valve A at 733/1 to "Inject" pulse (200 ms)
FILL B/STEP 1 ***001*******1 switches valve B at 733/1 to "Fill" pulse (200 ms)
INJECT B 1 ***110*******0 switches valve B at 733/1 to "Inject" pulse (200 ms)
ZERO 1 ***011*******0 triggers autozero at 732/1 pulse (200 ms)
PUMP 752 ON ************1* switches on pump at 752/753 static
PUMP 752 OFF ************0* switches off pump at 752/753 static
STEP MSM 753 ***********1** switches 753 suppressor module to next position pulse (200 ms)
[PUMP R/S 2 ] ******0*001*** starts/stops 709/2 IC Pump static
[FILL A 2 ] ******0*010*** switches valve A at 733/2 to "Fill" static
[INJECT A 2 ] ******1*000*** switches valve A at 733/2 to "Inject" static
[FILL B/STEP 2] ******0*101*** switches valve B at 733/2 to "Fill" static
[INJECT B 2 ] ******1*010*** switches valve B at 733/2 to "Inject" static
[ZERO 2 ] ******0*011*** triggers autozero at 732/2 static
[PUMP 754 ON ] *******1****** switches on pump at 754 (with 6.2143.220 cable) static
[PUMP 754 OFF ] *******0****** switches off pump at 754 (with 6.2143.220 cable) static
SCAN commands
Parameter Bit pattern Function
Ready1 *******1 waits until the "ready" state of 732/1 is reached
End1 ****1*** waits until the 732/1 sends the EOD signal
End2 *1****** waits until the 732/2 sends the EOD signal
Wait1 *****1** waits until the 732/1 sets the remote line 3 to 1
Wait2 ***1**** waits until the 732/2 sets the remote line 3 to 1
Wait* ***1*1** waits until the 732/1 and 732/2 set their remote line 3 to 1
Pump1 ? ******1* waits until the 709/1 IC Pump is running
Pump2 ? **1***** waits until the 709/2 IC Pump is running
Pump* ? **1***1* waits until the 709/1 and 709/2 IC Pumps are running