User Manual
4 Operation
766 IC Sample Processor
88
>final sequence
1 CTL:Rm: PUMP R/S 1
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------
− Switch off 709 IC Pump
− —— Settings for changer functions ——
− —— Reaction to manual stop ———
Settings for 732 IC Detector
Setting of measurement parameters and creation of a program ac-
cording to section 3.3.2 of the 754 Instruction for Use. For the program
type, select remote instead of cycle.
Settings in the «IC Metrodata» program
Definition of an IC method without method associated remote control.
The analysis time Duration must be ca 1 min shorter than the analysis
time entered in the 732 IC Detector.
Procedure at program start
1. Enter parameter number of samples for 766 IC Sample Processor.
2. Create new queue in the program «IC Metrodata» and start it.
3. Start 766 IC Sample Processor with <START>.