4.6 Standard methods
766 IC Sample Processor
87
>final sequence
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------
—— Settings for changer functions ——
—— Reaction to manual stop ———
Settings in the «IC Metrodata» program
Definition of IC method (analysis time, start parameters for IC instru-
ments 732, 733 and 709, etc.), additional association of this method
with the following time program for the 732 IC Detector:
Method / Equipment / Setup / Program / Text:
0.0 Puls_766_set Set remote line 3 at 732/1 to 1 (s. below)
0.1 Puls_766_reset Set remote line 3 at 732/1 to 0 (s. below)
0.2 Flag end Program end
Method / Equipment / Setup / Program / Remote configuration:
Puls_766_set = **1***** Set remote line 3 at 732/1 to 1
Puls_766_reset = **0***** Set remote line 3 at 732/1 to 0
Method / Equipment:
Start with method
Procedure at program start
1. Start 766 IC Sample Processor with <START>.
2. Create new queue in the program «IC Metrodata» and start it.
4.6.8 Method "Dialysis"
Application
IC determinations with automatic sample dialysis.
Interconnection
see Fig. 13 and section 2.3.9 as well as 754 Instructions for Use
Program of 766 IC Sample Processor
766 IC Sample Proc. 11111 5.766.0010
parameters
method Dialysis
number of sample: rack
>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Pump1 ?
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 CTL:Rm: PROG R/S 1
5 SCN:Rm : Ready1
Report header with serial number and program ver-
sion
Method name
Number of samples to be processed (entire sample
rack)
Initialize remote interface
Waiting until 709 IC Pump runs
Move needle to sample position
Place lift with needle to working position
Start time program at 732
Waiting until 732 IC Detector sends ready signal