User Manual

4 Operation
766 IC Sample Processor
86
4.6.7 Method "Preconc"
Application
Preconcentration of samples with very low concentrations in an pre-
concentration column (trace analysis).
Interconnection
see Fig. 13
Installation of the preconcentration column
The preconcentration column, consisting of the 6.1006.200 Metrosep
Anion preconcentration cartridge and the 6.2828.010 glass cartridge
holder, is installed at the injection valve of the 733 IC Separation Center
instead of the sample loop (see Fig. 18). It is important to install the
sample inlet from the 766 IC Sample Processor at connection 2 of the
injection valve in order to run the preconcentration column for the en-
richment and the sample injection in the opposite direction. If a larger
sample throughput is desired, install the optionally available 6.1826.040
pump tubing (inner diameter 1.6 mm, flow ca. 3 mL/min) at the peristal-
tic pump of the 766 IC Sample Processor.
12
4 5
3 6
12
4 5
3 6
Fig. 18: Installation of the preconcentration column
Program of 766 IC Sample Processor
766 IC Sample Proc. 11111 5.766.0010
parameters
method Preconc
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: INJECT A 1
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump1 ?
3 MOVE 1 : sample
4 LIFT: 1 : work mm
5 PUMP 1.1 : ON s
6 WAIT 60 s
7 CTL:Rm: FILL A 1
8 WAIT 120 s
9 PUMP 1.1 : OFF s
10 CTL:Rm: ZERO 1
11 CTL:Rm: INJECT A 1
Report header with serial number and program ver-
sion
Method name
Number of samples to be processed (entire sample
rack)
Initialize remote interface
Switch injection valve A at 733 to "Inject"
Waiting until 732 IC Det. sends signal on rem. line 3
Waiting until 709 IC Pump runs
Move needle to sample position
Place lift with needle to working position
Switch on pump at 766
Waiting time 60 s
Switch injection valve A at 733 to "Fill"
Sample preconcentration during 120 s
(preconcentration time must be adapted)
Switch off pump at 766
Trigger autozero at 732 IC Detector
Switch injection valve A at 733 to "Inject"
Waste
Sample
Sample
Waste
Sample injection
Preconcentration
Preconc. column
Eluent
IC column
Eluent