4.1 Fundamentals of operation
766 IC Sample Processor
39
Key Normal state Edit mode
Turn rack right
The <è> key turns the sample
rack one position to the right, i.e. in
the clockwise direction. The next
lowest beaker position is placed
under the needle.
The turning speed of the rack can
be defined in the Parameter menu.
The rotation of the rack can only be
carried out when the lift is at or
above the shift position.
Move the cursor one column
to the right
With the <è> key the cursor is
moved one column to the right in
an editing line with two parameters.
Any data which has been altered
will not be carried over during this
action (see <ENTER> key).
INSERT
Add a command line to a
sequence
Adds a new command line above
the current line in a sequence.
The NOP command (no operation)
automatically occupies this line and
has no function.
The lines following this line are
shifted one line downwards.
DELETE
Delete a command line in a
sequence
Deletes the current line in a
sequence.
The lines which follow shift upwards
by one line.
TOWER
SELECT
Select parameter
With the <SELECT> key given
data values can be selected for a
particular parameter in manual
operation.
With every repeated keystroke the
next value that can be selected is
displayed. The last value is followed
again by the first.
The data is accepted with
<ENTER>.