User Manual
4 Operation
766 IC Sample Processor
38
Key Normal state Edit mode
END
Lift in work position
• The <END> key runs the lift into
the work position.
• The work position is defined
separately for every sample rack in
the configuration menu under
'>rack definitions' (in mm from
the rest position, i.e. as measured
from the upper stop).
Select the last line of a menu
• With the <END> key, the last line
in a menu or a sequence can be
accessed.
• Any data that has been altered in a
menu or command line is not car-
ried over (see <ENTER> key).
Run lift upwards
• Run lift upwards. The lift movement
is executed as long as the key
remains pressed.
Select previous menu line
• In a Select menu or a sequence the
<
é
> key accesses the previous
line.
• Any data that has been altered in a
menu or command line is not car-
ried over (see <ENTER> key).
Run lift downwards
• Run lift downwards. The lift
movement is executed as long as
the key remains pressed.
Select next menu line
• In a Select Menu or a sequence the
<
ê
> key accesses the next line.
• Any data that has been altered in a
menu or command line is not car-
ried over (see <ENTER> key).
Turn rack left
• The <ç> key turns the sample
rack one position to the left, i.e. in
the counterclockwise direction. The
next highest beaker position is
placed under the needle.
• The turning speed of the rack can
be defined in the Parameter menu.
• The rotation of the rack can only be
carried out when the lift is at or
above the shift position.
Move the cursor one column
to the left
• With the <ç> key the cursor is
moved one column to the left in an
edit line with two parameters.
• Any altered data will not be carried
over during this action (see
<ENTER> key).