User Manual
3 Operating tutorial
766 IC Sample Processor
34
<QUIT>
<
ê
>
<ENTER>
>sample sequence
9 NOP
• An empty line with a 'NOP'-entry (no operation) always forms the end
of a sequence.
• Exit the sample sequence with <QUIT> and go to the final se-
quence.
START
• After all sample tubes have been processed, the final sequence is
executed.
>final sequence
1 CTL:Rm: PUMP R/S 1
• In this line the pump drive at the 709 IC Pump is switched off. Press
<START>.
START
>final sequence
2 CTL:Rm: PUMP 752 OFF
• In this line the pump drive at the 752 Pump Unit is switched off.
Press <START>.
QUIT
>final sequence
3 NOP
• Now you have reached the end of the final sequence and have
completed the entire run of a sample series.
• By pressing <QUIT> twice the 766 IC Sample Processor returns to
the normal state.
PARAM
7
SAMPLE
START
• Now prepare some sample tubes and place them on the sample
rack. Prepare the IC system with 732 IC Detector, 733 IC Separation
Center, 709 IC Pump and 752 Pump Unit for the recording of chro-
matograms.
• Enter the number of samples to be processed (<PARAM>) and
define the position of the first sample (SAMPLE = 1).
• Now you can start your first sample series with <START>.