User Manual

3 Operating tutorial
766 IC Sample Processor
28
3.2 Configuration
CONFIG
ENTER
TOWER
SELECT
The dialog language can be set in the configuration menu. Press
<CONFIG>
configuration
>auxiliaries
and then <ENTER>.
>auxiliaries
dialog: english
This menu item has a colon, indicating that here the parameters can
be selected from a list. Press <SELECT> several times in order to
view the various selections and get used to this type of dialog.
>auxiliaries
dialog: deutsch
With <ENTER> you can accept the suggestion 'dialog: eng-
lish'.
>auxiliaries
display contrast: 3
4 × <
ê
>
<ENTER>
By pressing <
ê
> 4 times you reach the menu selection 'max. lift
way'.
>auxiliaries
max. lift way 125 mm
Here the lowest allowable lift position for automatic and manual
operation can be set. For the 6.2041.430 sample rack with
6.2743.050 sample tubes, this limit value of 125 mm should not be
changed. Accept the value by pressing <ENTER>.
configuration
>rack definitions
QUIT
or
STOP
In order to put the 766 IC Sample Processor back into the initial
position, press <QUIT> or <STOP>.
******** counter 0/127
PUMP- ready
In the normal state, the method name and the sample counter
reading are displayed in the first line. The second line serves as
status line which displays the pump status and the changer status.