User Manual
2 Installation
766 IC Sample Processor
16
2.3.8 Tubing connections 766 – 733
For transferring the sample from the 766 IC Sample Processor to the
injection valve of the 733 IC Separation Center the following tubing
connections must be made:
Pump tubings are consumable material with a lifetime which depends
on the contact pressure (see section 6.3.3). This is why the tubing
cartridges should be raised completely by loosening snap-action lever
1717 on the right-hand side if the pump is to remain switched off for a
considerable length of time (the set contact pressure remains un-
changed).
The 6.1826.0X0 pump tubing is made of PVC and must not be used
for rinsing with solutions which contain acetone. In such cases, rinse
with different pump tubing or a different pump.
1
Install pump tubing
• Release tubing cartridge 1313 from holding clamp 1616 by
pressing down snap-action lever 1717 and remove from
mounting pin 1212 on the 766 IC Sample Processor (see Fig.
1).
• Press contact pressure lever 1515 on the tubing cartridge down
as far as it will go.
• Insert pump tubing 1010 in the tubing cartridge 1313 as shown in
Fig. 9. The black-black stopper 3636 must click into the corre-
sponding holder on the left-hand side of the tubing cartridge.
• Place the tubing cartridge 1313 on mounting pin 1212 and press
down on the right-hand side until snap-action lever 1717 clicks
into position on holding clamp 1616. Take care that no kinks are
formed in the pump tubing.
2
Connection needle – pump tubing
• Mount the PEEK compression fitting 66 on one end of the
PEEK capillary 88 and the PEEK compression fitting 3434 on the
other end.
• Screw the PEEK capillary 88 with the PEEK compression fitting
66 on to the PEEK compression fitting 55 already mounted on
needle holder 44 (see Fig. 6).
• Insert PEEK capillary 88 into the guide chain 99 (see Fig. 1 and
section 2.3.8).
• Screw a coupling 3535 on to the PEEK compression fitting 3434 at
the other end of PEEK capillary 88.
• Push coupling 3535 on to the inlet end of the pump tubing 1010
(see Fig. 6).